Cleaning method development and validation of Brimonidine Tartrate Residue from Brimonidine Tartrate API

Sayyad. Kousr Ali*, Tanmoy Mandal


 

 

To Develop the Precise, Linear and Accurate Cleaning method Validation for the Brimonidine Tartrate from Brimonide tartrate Active Pharmaceutical Ingredient as per International Conference on Harmonization guidelines by using High Performance Liquid Chromatography.

Material and Methods:  In this method uses a reverse phase C18 column make as inert sustain AQ C18 column (250*4.6mm,5µ), mobile phase-a as 0.1% ortho phosphoric acid and mobile phase-b as acetonitrile : methanol (50:50) v/v , flow rate as 1.2 ml/min, and detection of wavelength AS 248 nm, column temperature 40°C, injection volume 10µL and autosampler temperature 25°C by using Photo Diode Array Detector.

Results: In this method Brimonidine tartrate peak was eluted at 3.9 minutes. Brimonidine tartrate linear range was 0.5 µg/mL to7.5 µg/mL. In System suitability the % Relative standard deviation observed as 0.5. In swab and Rinse samples recovery was found between  97 to 103%. The limit of Quantification and limit of Detection range is found to be 0.5µg/mL and 0.15 µg/mL respectively.

Conclusion: An accurate, linear and precise Reverse phase High Performance Liquid Chromatography method was developed and validated as per International Conference on Harmonization guidelines. Hence this cleaning method validation is used for routine analysis of the residue samples in various pharmaceutical manufacturing areas. 


अस्वीकृति: इस सारांश का अनुवाद कृत्रिम बुद्धिमत्ता उपकरणों का उपयोग करके किया गया है और इसे अभी तक समीक्षा या सत्यापित नहीं किया गया है।

में अनुक्रमित

  • कैस
  • गूगल ज्ञानी
  • जे गेट खोलो
  • चीन राष्ट्रीय ज्ञान अवसंरचना (सीएनकेआई)
  • उद्धरण कारक
  • ब्रह्मांड IF
  • इलेक्ट्रॉनिक जर्नल्स लाइब्रेरी
  • रिसर्च जर्नल इंडेक्सिंग की निर्देशिका (डीआरजेआई)
  • गुप्त खोज इंजन लैब्स
  • आईसीएमजेई

और देखें

Flyer