अमूर्त

Determination of purity of 24 carat (5N gold) using inductive coupled plasma-optical emission spectrometer

N.Raman, D.Sathiya Narayanan


Hallmarking centers determine the purity of 22 carat usually by cupellation (Fire assay) method as per IS 1417. In this method they are using 24 carat gold as an additive and the purity of 24 carat gold is also certified. The determination of purity of 24 carat gold involves the estimation of impurities present in it using Inductive Coupled Plasma-Optical Emission Spectrometer (ICP-OES) quantitatively. The general impurities are Ag, Cu, Pd, Fe, Pb,Mg,As, Bi, Sn, Cr, Ni,Mn, Ru, Ir, Pt, Os, Rh, Cd, Zn and Te as per STD ASTM B- 562. Upon adding the impurities present in gold, rest is purity by difference. Based upon the results, gold can be classified into different grades which are 99.500, 99.950, 99.990 and 99.995 % fineness. This analytical technique is most important for Hallmarking centers for ascertaining their 24 caratGold.


में अनुक्रमित

  • कैस
  • गूगल ज्ञानी
  • जे गेट खोलो
  • चीन राष्ट्रीय ज्ञान अवसंरचना (सीएनकेआई)
  • उद्धरण कारक
  • ब्रह्मांड IF
  • रिसर्च जर्नल इंडेक्सिंग की निर्देशिका (डीआरजेआई)
  • गुप्त खोज इंजन लैब्स
  • आईसीएमजेई

और देखें

Flyer